Taxes included.
Measurement of flow and temperature at the same time.
Optimised for water, oils and air.
Clearly visible 4-digit LED display.
With two scalable analogue outputs.
The process connection can be rotated for optimum alignment
| Number of inputs and outputs |
Number of analogue outputs: 2 |
| Process connection |
threaded connection M18 x 1,5 internal thread |
| Special feature |
Gold-plated contacts |
| Media |
water; glycol solutions; air; oils |
| Note on media |
| low-viscosity oils with viscosity: ≤ 40 mm²/s (40 °C) |
| high-viscosity oils with viscosity: > 40 mm²/s (40 °C) |
|
| Medium temperature [°C] |
-20...90 |
| Pressure rating [bar] |
100 |
| Pressure rating [MPa] |
10 |
| MAWP (for applications according to CRN) [bar] |
100 |
| Operating voltage [V] |
18...30 DC |
| Current consumption [mA] |
< 100 |
| Protection class |
III |
| Reverse polarity protection |
yes |
| Power-on delay time [s] |
10 |
Inputs / outputs
| Number of inputs and outputs |
Number of analogue outputs: 2 |
Outputs
| Total number of outputs |
2 |
| Output signal |
analogue signal |
| Number of analogue outputs |
2 |
| Analogue current output [mA] |
4...20; (scalable) |
| Max. load [Ω] |
350 |
| Short-circuit protection |
yes |
| Type of short-circuit protection |
pulsed |
| Overload protection |
yes |
Measuring/setting range
| Probe length L [mm] |
45 |
| Operating mode |
relative; absolutely liquid; absolutely gaseous;(absolute: reference measurement recommended; Factory setting: relative) |
Temperature monitoring
| Measuring range [°C] |
-20...90 |
| Resolution [°C] |
0.2 |
| Analogue start point [°C] |
-20...76 |
| Analogue end point [°C] |
4...100 |
| In steps of [°C] |
0.2 |
Liquid media - absolute operating mode
| Setting range [m/s] |
0.04...3 |
| Greatest sensitivity [m/s] |
0.04...3 |
Liquid media - relative operating mode
| Setting range [m/s] |
0.04...6 |
| Greatest sensitivity [m/s] |
0.04...3 |
Gases - operating mode "absolute"
| Setting range [m/s] |
0...100 |
| Greatest sensitivity [m/s] |
30...100 |
Gases - operating mode "relative"
| Setting range [m/s] |
0...200 |
| Greatest sensitivity [m/s] |
30...100 |
Accuracy / deviations
| Temperature drift [cm/s x 1/K] |
0,003 m/s x 1/K (< 20 °C; > 70 °C) |
| Temperature gradient [K/min] |
100 |
Absolute operating mode
| Repeatability |
0,05 m/s;(water; flow velocity: 0,05...3 m/s) |
Relative operating mode
| Accuracy |
± (7 % MW + 2 % MEW);(for relative mode in the range of maximum sensitivity under the following conditions:; water: 20...70 °C; inlet length: 1.5 m; DN25 (DIN 2448); mounting position according to instructions; Accuracy can differ for other media and mounting positions.) |
| Repeatability |
0,05 m/s;(water; flow velocity: 0,05...3 m/s) |
Temperature monitoring
| Temperature drift |
± 0,005 K/°C |
| Accuracy [K] |
± 0,3 / ± 1;(water; flow velocity: 0,3...3 m/s/ air; flow velocity: > 10 m/s) |
Response times
| Response time [s] |
0.5;(T09; water; glycol: 0,8 s; air: 7 s; oil: 1,8 s; each T09) |
Temperature monitoring
| Dynamic response T05 / T09 [s] |
1,5 (T09);(water; flow velocity: 0,3...3 m/s) |
Software / programming
| Parameter setting options |
medium selection; Damping; Teach function; display can be rotated and switched off; standard unit of measurement; process value colour |
Operating conditions
| Ambient temperature [°C] |
-40...80 |
| Storage temperature [°C] |
-40...100 |
| Protection |
IP 65; IP 67 |
Tests / approvals
| EMC |
|
| Shock resistance |
| DIN EN 60068-2-27 |
50 g (11 ms) |
|
| Vibration resistance |
| DIN EN 60068-2-6 |
20 g (10...2000 Hz) |
|
| MTTF [years] |
180 |
| UL approval |
| UL Approval no. |
I004 |
| File number UL |
E174189 |
|
| Weight [g] |
281 |
| Materials |
stainless steel (1.4404 / 316L); stainless steel (1.4310 / 301); PBT-GF20; PBT-GF30 |
| Materials (wetted parts) |
stainless steel (1.4404 / 316L); Gasket: FKM |
| Process connection |
threaded connection M18 x 1,5 internal thread |
Displays / operating elements
| Display |
| Display unit |
6 x LED, green (%, m/s, l/min, m³/h, °C, 10³) |
| measured values |
alphanumeric display, red/green 4-digit |
|
| Remarks |
| MW = measured value |
| MEW = Final value of the measuring range |
|
| Pack quantity |
1 pcs. |
Electrical connection
| Connection |
Connector: 1 x M12; coding: A; Contacts: gold-plated |
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — SA5004-00 — 13.04.2023 —